Hoentzsch explosion proof Multi functional handheld unit with data logger for measuring flow rate,flow velocity, temperature, pressure and other variables in explosive atmospheres
Hoentzsch explosion proof Multi functional handheld unit with data logger for measuring flow rate,flow velocity, temperature, pressure and other variables in explosive atmospheres

Handheld Flowtherm ex

SKU: Flowtherm ex

Manufacturer: Hoentzsch

Multi functional handheld unit with data logger for measuring flow rate,flow velocity, temperature, pressure and other variables in explosive atmospheres

download pdf datasheet

Handheld flowtherm Ex - in combination with flow sensors FA, VA, TA, Pt100 and others

Flowthermdescription

  • This user-friendly unit, which was developed with the latest safety standards in mind, excels thanks to the diversity of connectable sensors (vane wheel, vortex, thermal, Pt100 and any other physical sensor with analog output) thus allowing for a wide range of applications.
  • Measuring applications with the flowtherm Ex are manifold: gas mass flow and flow rate in air, exhaust air, process gases, particle and condensate laden emissions, vehicle exhaust emissions, in aggressive gases with high working temperatures and, most importantly, in the hazardous area Zone 1, et al.
  • ATEX-approved for potentially explosive atmospheres:  II 2G Ex ib IIC T4
  • That is why the new handheld flowtherm Ex lends itself especially for measuring on gas blow-outs and in gassing of utility gas pipelines, in hazardous areas of all industrial branches, municipal utilities and public authorities as well as for research and development.

    flowtherm handheld Ex display description Hoentzsch Flow measurement datalogger gas, liquids

  • Measurements with up to three sensors can be displayed simultaneously.
  • The data logger allows storage of up to 40,000 values with date and time.
  • Up to 100 different profiles can be stored for user or measuring points.
  • Various measurement modes (also long-term measurement).
  • Thanks to the USB interface, the logged data can be downloaded to a PC for configuration.
  • The flowtherm Ex has the following inputs: flow sensors (FA, TA, VA); Pt100; 4-20 mA; 0-10 V.
  • A 12 V supply is available for sensors in 4-wire system.
  • Mains or battery operation.

Specifications

  • Displayed indicators:
    • actual flow velocity [m/s], [ft/min]
    • actual flow rate [m³/h], ...
    • standard flow velocity [N-m/s], ...
    • standard flow rate [N-m³/h], ...
    • mass flow [kg/h]
    • temperature [°C], [K]
  • IP protection type IP65 (housing)
  • Type of protection ATEX category 2G (zone 1)
  • Sensor input FA; FT; FAR; VA; VAT; TA; TAT; Pt100; analog input 4 - 20 mA or 0 - 10 V